Femoral loads measured by a six-axis load cell.

Auteur(s)
Cheng, R. Denton, R.A. & King, A.I.
Jaar
Samenvatting

A six-axis femoral load cell was designed, fabricated, calibrated and tested in an ATD upper leg. It has linear characteristics with adequate output and dynamic response. Its output compared well with that of load cells mounted behind a knee bolster which was impacted by an ATD instrumented with this load cell. Although the axial load was found to be well below the present limit of FMVSS208, the bending moments were found to be high. A conservative tolerance limit for axial load should be considered and cadaver data are required.(a) the covering abstract for the proceedings is IRRD abstract no 254690.

Publicatie aanvragen

9 + 6 =
Los deze eenvoudige rekenoefening op en voer het resultaat in. Bijvoorbeeld: voor 1+3, voer 4 in.

Publicatie

Bibliotheeknummer
C 51428 (In: B 16296 [electronic version only]) /84 / IRRD 254779
Uitgave

In: Proceedings of the 23rd Stapp Car Crash Conference, San Diego, October 17-19, 1979, SAE Technical Paper 791012, p. 263-288, 10 ref.